Sanyou B-2000 Infant Incubator

STANDARD FEATURES:
●Servo Control Mode: air temperature, skintemperature;
● Parameter Displaying: set temperature,air temperature, skin temperature, internal humidity, heat power rate;
● Audible & Visual Alarms: powerfailure, sensor failure, higher & lower temp, air & skin temperature override,fan failure, system failure;
● >37℃Preset Mode: power failure reset memory, alarm silence, keypad lock, self-testfunction;
● Hood with dual slope; 6 hand-ports;
● RS-232 Output Connector (use for networksystem);
●IV poles;
●Protective guard rail;
●Storage cabinets.

 


LOADING IMAGES

Specifications of hood

Mattress to Hood: 450 mm
Observation side of hood: Two incline
Operate port: 6 (2 Iris ports)
Tubing grommets: 4
Front access panel: 826×310 mm
Mattress Tray Size: 630×360 mm

Technical Specifications:

Powerrequirement ≤400VA
Temperaturecontrol mode: Air and skin servo control
AirTemp Control Range: 20℃~37℃
Temperatureoverride mode to: 20℃~39℃
SkinTemp Control Range: 30℃~37℃
Temperatureoverride mode to: 30℃~38℃
Skintemp sensor accuracy: ≤0.3℃
Temperaturefluctuation: ≤0.5℃
Mattresstemperature uniformity: ≤0.8℃
Temperaturerise time: ≤45 min
Noiselevel within hood environment: ≤55dB

Alarms

Power failure alarm
Fan failure alarm
Sensor failure alarm
High air temperature alarm: +3.0℃
Low airtemperature alarm: -3.0℃
High skin temperature alarm: +1.0℃
Low skin temperature alarm: -1.0℃
Over air temperature alarm: ≤38.0℃; ≤40.0℃
Over skin temperature alarm: ≤38.0℃; ≤39.0℃
Systemfailure alarm
Alarmsilence function

Operational condition

Environment temperature: +20℃~+30℃
Environment relative humidity: 5%~99%
Environment air velocity of flow: <0.3m/s

Others

Infant bed tilt adjusting ±10° step-less
Mattress tray Max. load 10kg
Transfusion pole Max. load 2kg
CabinetMax. load 6kg