- Light source: 6V 10W, halogen lamp with more than 2000 hours’ life time
- Spectrometer: Interference filter
- Wavelength: 340nm, 380nm, 405nm, 505nm, 546nm, 578nm, 620nm, and one free position
- Flow cell: Permanent quartz flow cell, minimal volume is 32μL, and disposable cuvette is selectable
- Test method: End point, Fixed time, Kinetic, Factor, Multi-standards, Bichromatic, Dual wavelengths
- Screen: Touch screen
- Stray light: ≤0.5%(absorbance≥2.5)
- Absorbance range: 0 ~3.000
- Absorbance linearity: ±1%
- Absorbance repeatability: ±0.001
- Absorbance resolution: 0.0001
- Precision: ≤1.0%
- Carry over: ≤1%
- Temperature: 25℃, 30℃, 37℃, ±0.3℃
- Aspirationvolume: 200μL~3000μL
- Sample volume: 10μL~50μL
- Reaction curve: Real time display the reaction curve
- Memory: Able to store more than 200 parameters while 55 parameters of them are stablized; Able to save more than15000 test results and auto save the results if the power is off during working
- QC: Auto draw the QC figure
- Reagent: Open reagent
- Display: LCD with high brightness
- Input: Key board and touch screen
- Printer: Inner thermal printer, support on-line printing
- Input and output: RS-232 serial port, free software to connect with LIS
- WorkingCondition: Temperature: 10℃~30℃; Relative Humidity: 20%~80%
- Voltage: AC100-240V ± 10%, auto adapt according to the input voltage
- Power: ≤100W
- Quality Commitment: CE , ISO ,CMD.
PARAMETER
Test item | Parameter |
---|---|
Liver | ALT、ALP、GGT、TP、Alb、TBIL、DBIL |
Heart | LDH、CK、CK-MB、AST、 a-HBDH |
Lipids | CHO、TG、HDL_C、LDL_C |
Lipoprotein | Apo A1、Apo B、Lp(a) |
Kidney | BUN、CRE、UA |
Ion | Na、Mg、Ca、K、P、Cl |
Other | CO2-CP、AMS、LIP、GLU、IgA、IgG、IgM、C3、C4、CRP |